A
simple, low-cost millimeter-wave (70GHz) interferometer with
a phase accuracy better than +/- 2 degrees, and a response time
of 10ns has been developed and implemented as a diagnostic tool
at the LAPD. The simplicity of this interferometer makes it
ideal for measurement of electron densities in a laboratory
plasma. The relative high accuracy and low cost are attained
by a homodyne system with a quadrature intermediate frequency
(IF) mixer, now commercially available in millimeter-wave frequency
bands. Since the system isolator operates near a magnetically-confined
plasma, it has been protected by magnetic shielding.
Click
here to read A
Reliable Millimeter-Wave Quadrature Interferometer by M.
Gilmore, W. Gekelman, K. Reiling
This
is an unpublished internal lab paper.
Feel free to copy and use it as you wish.